Figure 1: Clinical applications of Zexira i9 at our hospital.
Figure 2: Compact design of Zexira i9.
Figure 3: Wide range of movements and image acquisition modes maximizes access to the target region
Figure 4: Variable SID (110 cm to 180 cm)
Figure 5: Use of a long SID in videofluorographic swallowing studies Movement of the contrast medium can be clearly observed. (pulse rate: 15 fps, dose mode: Normal)
Figure 6: i-fluoro (function that allows the desired region to be observed without moving the tabletop or the imaging system)
Figure 7: Use of i-fluoro, in which the field of view is moved but the tabletop and imaging system do not move
Figure 8: Example of use of i-fluoro during nephrostomy catheter replacement in the right kidney
Figure 8: Example of use of i-fluoro during nephrostomy catheter replacement in the right kidney
Figure 9: Example of use of i-fluoro for contrast-enhanced imaging during vesicostomy (pulse rate: 7.5 fps, dose mode: Normal)
Figure 9: Example of use of i-fluoro for contrast-enhanced imaging during vesicostomy (pulse rate: 7.5 fps, dose mode: Normal)
Figure 10: Stitching radiography of X-ray protective equipment
Figure 11: Tomosynthesis
Figure 12: Dose reduction due to a larger number of pulse rate settings etc.
Figure 13: Pediatric VCG (pulse rate: 3 fps, dose mode: Low)
Figure 14: Nerve root block (pulse rate: 7.5 fps, dose mode: Mid)
Figure 14: Nerve root block (pulse rate: 7.5 fps, dose mode: Mid)
Figure 15: Dose report function of Zexira i9
Figure 15: Dose report function of Zexira i9
© CANON MEDICAL SYSTEMS MALAYSIA SDN. BHD.
© CANON MEDICAL SYSTEMS MALAYSIA SDN. BHD.